Electrode
calibration not
required

Message appears at the start of a sequence, if a method sequence
has been programmed with an electrode calibration. The electrode
used has been programmed without calibration Calibration
request = No.

Go to Sequence/Sample stack, Edit sequence menu and
remove the electrode/reagent calibration method.

Electrode
calibration
parameters

This menu contains the general parameters concerning the
pH electrode calibration method (measurement stabilisation criteria
in particular).

To access:

1.

From the Electrode window, press 4.

2.

Select the electrode to be edited.

3.

Press 2 Edit electrode and check that the Calibration
request = Yes option has been selected.

4.

Edit the electrode calibration general parameters.

5.

Use the LEFT/RIGHT arrow keys to move to the last Edit
electrode display.

6.

Press 1 Calibration parameters.