Electrode calibration not required, Electrode calibration parameters, See "electrode calibration – Hach-Lange TITRALAB 860_865 Reference Manual User Manual
Page 100: Titra lab

Titra
Lab
®
TitraLab 860 and 865 Reference Manual
Page 100
Electrode
calibration not
required
Message appears at the start of a sequence, if a method sequence
has been programmed with an electrode calibration. The electrode
used has been programmed without calibration Calibration
request = No.
Go to Sequence/Sample stack, Edit sequence menu and
remove the electrode/reagent calibration method.
Electrode
calibration
parameters
This menu contains the general parameters concerning the
pH electrode calibration method (measurement stabilisation criteria
in particular).
To access:
1.
From the Electrode window, press 4.
2.
Select the electrode to be edited.
3.
Press 2 Edit electrode and check that the Calibration
request = Yes option has been selected.
4.
Edit the electrode calibration general parameters.
5.
Use the LEFT/RIGHT arrow keys to move to the last Edit
electrode display.
6.
Press 1 Calibration parameters.