Sample unit
Unit of the sample amount. The sample unit can be expressed in
weight (g, mg) or volume (ml, µl). The sample unit indicates the result
unit available.
Enter in:
Edit method > Sample menu
Sample unit
conflict
In a Coupled method, the sample unit is not the same for the two titra-
tion methods.
Check the Sample screens of each method of the Coupled method.
Select electrode
Routine user
Is able to select an electrode to check the parameters and/or start a
calibration.
1.
Select the method/sequence using the electrode.
2.
Press 1 to start a calibration or 3 to check the electrode parame-
ters. In both cases, the electrodes available will be those specified
in the actual working method or sequence.
Supervisor
From the Electrode window.
1.
Press 4.
2.
In the ID field, select the electrode from the User list.
Select method
To select a single method:
1.
Select Working mode = Method in the Main window.
2.
Press 3.
3.
In the ID field, select the method from the User list.
To select a method to be run using a Sample Changer:
1.
Select Working mode = SAC Method in the Main win-
dow. If this option is not available, enter the Setup menu > Config-
uration and define a sample changer.
2.
Press 3.
3.
In the ID field, select the method from the User list.