Monotonic IP
method
For this type of method, a preset volume of titrant (increment) is
added to the solution as soon as the measurement variations satisfy a
stability criterion or at the end of a preset acceptation time.
Figure 26: Monotonic IP method - size of the increments
In Monotonic IP, one measurement is saved per increment added.
Select a Monotonic IP method when you want to minimise the number
of increments added and in some conditions to speed up a sample
titration.
The Monotonic IP method is stopped after a given number (1 to 8) of
equivalent points has been detected.
Nb lines per
page (printouts)
When printing page by page (
see "Format (printouts)", page 119
),
this parameter sets the maximum number of lines for one printed
page.
Access:
Setup Menu > Configuration
Range available:
26 to 255
No curve
generation for a
reprocessing
Automatic mode reprocessing error: In the Coupled method, no
method is generating a curve before the method used to reprocess a
curve in automatic mode.
Review the Coupled method programmation.
Refer to "Reprocessing a curve (Automatic mode)", page 181
.
E or pH