Working in Routine mode

A.Access methods

A Routine operator has access to all the methods “Select method” and programmed parame-
ters “Display method” for checking purposes

B.Running methods

When working in “ROUTINE” mode, it is necessary to install your titration system according to
the selected method or sequence, prior to running a method or sequence.

1. Select the method or sequence

Refer to "Select method", page 203

.

Refer to "Select sequence", page 204

.

2. Check icons

Refer to "Check icons", page 19

.

Depending on the icons displayed, the Titration Manager will automatically guide you through
the steps necessary to run the analysis, see below:

a. Connect the electrode(s)

Connect/install electrodes and temperature sensors,

Refer to "Electrode connection",

page 102

.

b. Install regents(s)

Check that the burette is installed, if not,

see "Install burette", page 125

. Now, install

the reagent,

see "Install reagent", page 126

.

c. Calibrate electrode(s)

Now, run the calibration,

see "Calibrate electrodes", page 70

.

d. Calibrate reagent(s) or Enter titre

Now, run the calibration or enter the titre.

Refer to "Reagent calibration", page 168

.

Refer to "Enter titre", page 109

.

e. Run the method or the sequence

Refer to "Running a method", page 216

.

Refer to "Running a TIM sequence", page 218

.

Refer to "Running a SAC sequence", page 217

.