Titra lab, Working in routine mode – Hach-Lange TITRALAB 860_865 Reference Manual User Manual
Page 21

Titra
Lab
®
TitraLab 860 and 865 Reference Manual
Page 21
Working in Routine mode
A.Access methods
A Routine operator has access to all the methods “Select method” and programmed parame-
ters “Display method” for checking purposes
B.Running methods
When working in “ROUTINE” mode, it is necessary to install your titration system according to
the selected method or sequence, prior to running a method or sequence.
1. Select the method or sequence
Refer to "Select method", page 203
Refer to "Select sequence", page 204
2. Check icons
Refer to "Check icons", page 19
.
Depending on the icons displayed, the Titration Manager will automatically guide you through
the steps necessary to run the analysis, see below:
a. Connect the electrode(s)
Connect/install electrodes and temperature sensors,
Refer to "Electrode connection",
.
b. Install regents(s)
Check that the burette is installed, if not,
see "Install burette", page 125
. Now, install
see "Install reagent", page 126
.
c. Calibrate electrode(s)
Now, run the calibration,
see "Calibrate electrodes", page 70
d. Calibrate reagent(s) or Enter titre
Now, run the calibration or enter the titre.
Refer to "Reagent calibration", page 168
Refer to "Enter titre", page 109
.
e. Run the method or the sequence
Refer to "Running a method", page 216
.
Refer to "Running a TIM sequence", page 218
.
Refer to "Running a SAC sequence", page 217
.