Demand: locked, Demand: unlocked, Derivative – Hach-Lange TITRALAB 860_865 Reference Manual User Manual

Page 86: Titra lab

Advertising
background image

Titra

Lab

®

TitraLab 860 and 865 Reference Manual

Page 86

Demand:
Locked

Electrode or Reagent calibration

The routine user is not allowed to bypass an electrode and/or reagent
calibration demand before continuing measurements. It means that
the electrode and/or reagent calibration periodicity has(have) been
elapsed.

QC sample analysis

If a QC sample periodicity has been reached, the next run of the
method must be performed on a QC sample.

Sequence edition

The routine user is not allowed to create, edit or delete sequence of
methods.

Enter in:

Setup menu > Access routine mode

Demand:
Unlocked

Electrode or Reagent calibration

The routine user is allowed to bypass an electrode and/or reagent cal-
ibration demand and continue measurements. This happens when the
electrode or the reagent calibration periodicity has elapsed.

QC sample analysis

If a QC sample periodicity has been reached, the routine user is able
to run the method without having to use a QC sample.

Sequence edition

The routine user is allowed to create, edit or delete sequence of
methods.

Enter in:

Setup menu > Access routine mode

Derivative

Specify if you want to print the derivative curve at the end of each test.

Enter in:

Printouts menu of Monotonic IP, Dynamic IP or Continuous IP method.