PC keyboard
Specify the PC keyboard in use. For example, English (US) for a
Qwerty keyboard.
Enter in:
Setup menu > Configuration
Range available:
English (US), French, German, Spanish, Italian, Danish, Swedish.
Periodicity
Available if Electrode calibration request = Yes or Titre = Calibrate
Maximum period of time between two calibrations. If the period of time
is exceed measurements can no longer be performed using this elec-
trode or reagent. A new calibration is required except if Alarm :
Unlocked has been set in the SETUP > Access Routine mode
menu.
Enter in:
Edit electrode menu
Edit reagent menu
Range available:
1 to 999 days
Periodicity for
QC samples
This is the number of samples to be placed between two successive
QC samples. When this number is reached, a QC analysis must be
run using this method except if Alarm : Unlocked has been
set in the SETUP > Access Routine mode menu.
Enter in:
QC data menu
Range available:
1 to 999 samples