Programming electrodes

1.

Press 4.

2.

Press 1.

3.

Select function.
Select ID.

4.

Select ID from Catalogue or
Others list.
Press 1 to confirm.

5.

Press 1 to confirm the creation
of the new electrode.

6.

For a combined or a simple
electrode or for a reference
electrode, enter the potential (in
mV) of the reference versus the
Standard Hydrogen Electrode
(SHE).
For a combined or a simple
electrode if you have selected
the Others list, enter the inter-
nal pH of the electrode.
Enter the address of the elec-
trode.
Select Yes if a calibration is re-
quired, go to step 7.
Select No, for no calibration,
press Esc to leave the menu.
Programming is completed.