Coupled method, Titra lab – Hach-Lange TITRALAB 860_865 Reference Manual User Manual

Page 81

Advertising
background image

Titra

Lab

®

TitraLab 860 and 865 Reference Manual

Page 81

Coupled
method

A Coupled method is a combination of methods performed in the
same beaker. When using a coupled method, the instrument runs all
these methods on the same sample.

If you wish to run a series of methods in different beakers, it is neces-
sary to program a Sequence instead of a Coupled method.

Example: Combination of method 1 and method 2.

The number of test portions (for example 3) is entered during pro-
gramming. The method is then repeated in the number of beakers
specified.

Figure 15: Coupled method with three tests

= beaker number 1

Sample 1

1

1

1

2

2

3

3

Test portion 1

Test portion 2

Test portion 3

Method 1

Method 2

Method 1

Method 2

Method 1

Method 2

Advertising