Addition, Titra lab – Hach-Lange TITRALAB 860_865 Reference Manual User Manual
Page 47

Titra
Lab
®
TitraLab 860 and 865 Reference Manual
Page 47
Addition
Option available with TitraLab 865 or TitraLab860 with one or two
ABU52 connected.
Select Yes to carry out a reagent addition at the start of the titration or
a reagent calibration. This addition takes place before the titration
starts (predose included).
The reagent is added using a second burette controlled by the Titration
Manager. This second burette can be installed on the Titration
Manager or on an ABU52 connected to the Titration Manager,
Edit an Addition method when you want to perform 2 or 3
reagent additions successively or simultaneously.
Refer to "Addition method - definition", page 49
Use a sample preparation when you want to add a solvent
using a pump installed on a SAC850 or SAC950 Sample
Changer.
Refer to "Sample preparation no.", page 201
.
Enter in:
Edit method > Parameters
Edit reagent > Calibration parameters