TitraLab 860 and 865 Reference Manual

Page 87

Detailed

This parameter sets level of details of report printouts.

Detailed = Low

•

The header only comprises the analysis name, time and date and
the instrument serial number. These data are printed on the same
line.

•

Reagent or electrode calibration method: results are printed.

Detailed = Medium

This is the printout level selected by default.

•

The header comprises the analysis name, time and date, the
instrument serial number and the laboratory coordinates.

•

Reagent or electrode calibration method: results are printed.

Detailed = High

This is the printout level selected by default.

•

The header comprises the analysis name, time and date, the
instrument serial number and the laboratory coordinates.

•

Electrode data, electrode serial number, electrode calibration data
and results, burette serial number, reagent data and reagent
calibration data and results are printed.

•

The buffer data (name and batch number, potential value) are
printed.

•

End point/Inflection point coordinates (potential, volume) are
printed (except for a reagent calibration).

Enter in:

Edit method > Printouts
Edit reagent > Printouts
Edit electrode > Printouts

Dilution (Yes/
No)

Select Dilution = Yes when you are analysing samples which
have been diluted. The Final dil. amount and Aliquot
are to be entered next.

Enter in:

Edit method > Sample menu