TitraLab 860 and 865 Reference Manual

Page 87

Detailed

This parameter sets level of details of report printouts.

Detailed = Low

The header only comprises the analysis name, time and date and
the instrument serial number. These data are printed on the same
line.

Reagent or electrode calibration method: results are printed.

Detailed = Medium

This is the printout level selected by default.

The header comprises the analysis name, time and date, the
instrument serial number and the laboratory coordinates.

Reagent or electrode calibration method: results are printed.

Detailed = High

This is the printout level selected by default.

The header comprises the analysis name, time and date, the
instrument serial number and the laboratory coordinates.

Electrode data, electrode serial number, electrode calibration data
and results, burette serial number, reagent data and reagent
calibration data and results are printed.

The buffer data (name and batch number, potential value) are
printed.

End point/Inflection point coordinates (potential, volume) are
printed (except for a reagent calibration).

Enter in:

Edit method > Printouts
Edit reagent > Printouts
Edit electrode > Printouts

Dilution (Yes/
No)

Select Dilution = Yes when you are analysing samples which
have been diluted. The Final dil. amount and Aliquot
are to be entered next.

Enter in:

Edit method > Sample menu