User’s rights

The Supervisor and Routine modes set the user’s rights as shown in
the table below:

Figure 36: User’s rights

Viewing global
variables

1.

Enter the Main window:

2.

Press 5 GLP-Archives,

3.

Press 2 Global variable

4.

Scroll the G1 to G10 global variables using the arrow keys.

Supervisor

Routine

Create, edit sequence.
Remove all methods from a sequence.
Delete a sequence

Yes

Yes (if the
Supervisor
gives the
right to),

see
"Demand:
Unlocked",
page 86

.

Create method, electrode, reagent

Yes

No

Copy method, electrode, reagent

Yes

No

Program sequence, method, elec-
trode, reagent

Yes

No

Delete method, electrode, reagent

Yes

No

Reset method, electrode, reagent
parameters

Yes

No

Reset parameters to factory settings

Yes

No

Select sequence, method, electrode,
reagent

Yes

Yes

Check method, electrode, reagent
parameters

Yes

Yes

Install/replace/remove reagents

Yes

Yes

Connect/replace/disconnect elec-
trodes

Yes

Yes

Run sequence, method, electrode and
reagent calibrations

Yes

Yes

Consult results

Yes

Yes