Blank volume, Burette functions menu, Burette speed – Hach-Lange TITRALAB 860_865 Reference Manual User Manual

Page 69: Titra lab

Advertising
background image

Titra

Lab

®

TitraLab 860 and 865 Reference Manual

Page 69

Blank volume

Volume of titrant consumed by the titration on a blank sample.

If you run an End point titration with several E.P., a blank volume is
found for each E.P. If you run an Inflection point titration, one blank
volume is found irrespective of the number of Inflection points found
for the method. This blank volume is determined on a particular inflec
tion point of the titration curve,

see "Min. ordinate Max. ordinate -

Blank", page 137

.

When you run a method with a blank for the first time, the blank is
determined and stored along with the method. If the method is
restarted, you will be given the choice to run a sample analysis or a
blank analysis.

For a sample analysis using an End point method, a blank volume will
be automatically subtracted from each end point found.
For a sample analysis using an Inflection point method, a blank
volume will be automatically subtracted from the Inflection point
selected by the user,

see "Blank on inflection no.", page 68

.

Burette
functions menu

These functions can be activated during the preparation of the
burette, before installing the reagent.
To access, press 7 in the Reagents window. To run a burette function,
select the burette in the Address field, then press 1 to 6 corre-
sponding to the required procedure. The title bar shows the volume of
the burette in use. You can replace (key 5) or remove this burette (key
6).

Figure 12: Burette functions menu (burette installed/not installed).

Burette speed

The maximum burette speed (in ml/min) is three times the nominal
value of the burette per minute. However, for the 50 ml burette the
maximum burette speed is 1.5 times the nominal value, i.e. 75 ml/min.

Advertising