Ph0(25), Ph buffer, Ph calibration results parameters – Hach-Lange TITRALAB 860_865 Reference Manual User Manual

Page 149: Ph calibration solutions parameters, Titra lab

Advertising
background image

Titra

Lab

®

TitraLab 860 and 865 Reference Manual

Page 149

pH0(25)

This is the pH of the solution at 25°C at which the measured electrode
potential is equal to zero. The potential developed depends on both
electrodes; the reference and the glass. These potentials may vary to
bring about an electrode drift. This drift can be compensated by fre-
quent calibrations.

Normally the following ranges are used by default (6.850 pH and
7.200 pH)

Enter in:

Edit electrode > Results

Range available:

pH(0)25 min. = 0.000 pH to pH0(25) max.

pH0(25) max. = pH0(25) min. to 14.000 pH

pH buffer

Refer to "Solution menu", page 210

.

pH calibration
results
parameters

Refer to "Results menu", page 188

.

pH calibration
solutions
parameters

Refer to "Solution menu", page 210

.