Acceptance
criteria
Acceptance criteria = Yes
Enables the user to enter preset minimum and maximum values for
measurement results. If the result lies outside these values an alarm
message appears and the results are rejected by the instrument. The
Supervisor is the only person allowed to accept a result that has been
rejected by the instrument,
see "Result accepted (Yes/No)", page
Therefore, acceptance limits can be set on:
•
the result value such as a pH, a potential,
see "Minimum value - Maximum value", page 139
•
the response slope of a pH electrode,
see "Min. sensitivity - Max. sensitivity", page 137
.
•
see "Min. pH0(25) - Max. pH0(25)", page 137
Acceptance criteria = No
The Supervisor or Routine user is free to accept/reject the results.
Enter in:
Edit method > Results
Edit method > QC data
Edit reagent > Results
Edit electrode > Results
Irrespective of the Yes or No option selected for the
Acceptance criteria parameter, acceptance limits must be set
for the sample or the standard measured temperature,
see "Min. Temp. - Max. Temp.", page 138
,
see "T°C minimum/maximum value", page 219
.