Min. ordinate max. ordinate, Titra lab – Hach-Lange TITRALAB 860_865 Reference Manual User Manual

Page 136

Advertising
background image

Titra

Lab

®

TitraLab 860 and 865 Reference Manual

Page 136

Min. ordinate
Max. ordinate

Set the lower and upper limits of a search interval for a given inflection
point (IP). Min. and Max. ordinates are entered for each IP to be deter-
mined.

Figure 25: Minimum and maximum ordinates

Only one inflection point (IP) is determined for each Min./Max. ordinate
interval.

When several IP are found in the same interval:

The first IP is the result,

The other IP are rejected and a note “IP>1” is displayed.

If you know in advance the pH or mV of the inflection point, define a
search interval corresponding to the expected value.

Enter in:

Parameters menu of a Monotonic IP, Dynamic IP or Continuous IP
method.

Range:

Min. ordinate: -9.000 pH (or 2000 mV) to Max. ordinate
Max. ordinate: MIn. ordinate to +23.000 pH (or -2000 mV).

Advertising