Equation
formula

One or two user defined equations can be entered to calculate results
using user required units.

The formula specifies how the calculations should be performed. A
formula can use up to 64 characters and can consist of operators and
operands, that are derived from the completed titration data.

Enter in:

Edit method > Result

Table 5: Equation formula: operands

Operands

1.23, -24, -2.34E-5

Constants

pi

3,141592654

e

2,718281828 (base of the Natural loga-
rithm)

Ri where i = 1 to 8

Result no. i calculated for the method

Ei where i = 1 to 4
(for EP) and 1 to 8
(for IP)

Potential difference or pH measured at
EP/IP no. i

Vi where i = 1 to 4
(for EP) and 1 to 8
(for IP)

Cumulate volume delivered at EP/IP no. i

V

Total volume delivered

SA

Sample amount

DA

Final dilution amount

AL

Aliquot

T

Last temperature measured (in °C)

CT

Titrant concentration

CE

Excess reagent concentration

VE

Excess reagent volume

Bi with i = 1 to 4 (for
E.P) i = 1 (for I.P)

Blank volume for EP/IP no. i

Gi with i = 1 to 10

Global variables

F

Results factor