Proportional
band

This is the pH (mV) span prior to the end point, at which the titrant
speed will be reduced. The speed will increasingly reduce as the end
point is approached.

For titrations where the slope of the curve is high near the end point,
the proportional band setting should be large. If overshooting of the
end point frequently occurs, increase the Proportional band
or decrease the Max. speed setting. A high value for the Pro-
portional band will result in slow titrations.

Enter in:

Edit method > Parameters menu (end point method)
Edit reagent > Calibration parameters menu (end point method)

Range available:

-9.000 pH to +23.000 pH or -2000.0 mV to +2000.0 mV

QC (result
indicator)

Refer to "Result indicators", page 185

.

QC analysis
required

This message is displayed at the start of a method requiring a QC
sample.

Press

✓ and run a QC analysis.

End point no. n

Maximum
speed

Minimum
speed

Titrant delivery speed

mV or pH