Result id, Result indicators, Titra lab – Hach-Lange TITRALAB 860_865 Reference Manual User Manual
Page 185

Titra
Lab
®
TitraLab 860 and 865 Reference Manual
Page 185
Result ID
The name assigned by default to a method result number n is Rn (n=1
to 8). You can change the Rn name of each method result (16
alphanumerical characters maximum).
Enter in:
Edit method > Results
Edit electrode > Results
Edit reagent > Results
Refer to "Results menu", page 188
If you change a method result name (for example, from R1 to
«My result»), the equations taking this result into account
continue to use the original operand name (R1 in our
example).
Refer to "Equation formula", page 110
Result
indicators
Appear with the result at the end of a run.
•
OK: accepted by the instrument (result lies within the acceptance
limits).
•
Low: alarm, rejected by the instrument (result lies below the
acceptance limit).
•
High: alarm, rejected by the instrument (result lies above the
acceptance limit).
•
Time max: the measurement has been accepted at the end of the
Acceptance delay.
•
QC: the user has bypassed a QC sample analysis demand.
•
Man. stop: result has been obtained from a titration that was
stopped manually (Del key).
•
0IP: no inflection points found.
•
IP>1: more than one inflection point has been found in the min/max
ordinate window,
see "Min. ordinate Max. ordinate", page 136
You can reject a result that was accepted by the instrument.
You can accept a result that was rejected by the instrument
(Supervisor users only).
Refer to "Result accepted (Yes/No)", page 184
.