Back titration no/manual/ automatic, Back titration no/manual/automatic, Titra lab – Hach-Lange TITRALAB 860_865 Reference Manual User Manual

Page 58

Advertising
background image

Titra

Lab

®

TitraLab 860 and 865 Reference Manual

Page 58

Back titration
No/Manual/
Automatic

Back titration = No

This is a direct titration.

Back titration = Manual

This is a back titration where the excess reagent is added using a
separate burette or a pipette not controlled by the Titration Manager.
This reagent is defined using its titre and volume in the Parameters
screen. This reagent is not selected in the Reagent library.

Back titration = Automatic

Option available with TitraLab 865 or TitraLab860 with one or two
ABU52 connected.

This is a back titration where the excess reagent is added using a sec-
ond burette controlled by the Titration Manager. This second burette
can be installed on the Titration Manager or on an ABU52 connected
to the Titration Manager,

see "ABU1/ABU2", page 43

.

The excess reagent is created or selected from the User list.

Enter in:

Edit method > Parameters menu

Advertising