Cell grounding

Defines the grounding of the measuring cell. Select one of the follow-
ing options:

Reference

Grounding is ensured by a reference electrode - general use.

Metal

Grounding is ensured by a metal electrode connected to the GND
socket on the Titration Manager. Use this option in case of high resis-
tive solutions in order to avoid measuring background noise at the
electrodes.

Others

Grounding is not ensured by the reference electrode and is defined
outside the method.

Enter in:

Edit method menu

Cell window

Use LEFT/RIGHT arrow keys to access.
This window controls the stirring function of the measurement cell.

An external stirrer (propeller) can be connected to a Titration Manager
or an ABU52.

Refer to "Stirring", page 215

.

Change
electrode name

1.

Display the Electrode window.

2.

Press 4 then 2.

3.

In the ID field, enter the new name for the electrode (16 charac-
ters maximum).

Start/stop stirrer

Select stirring speed: 100 to 1100 rpm

Animated icon: indicates when stirrer

or propeller is in operation

Select the instrument having the stirrer

(TIM for Titration Manager, ABU1 or ABU2

for an ABU52).