Cell grounding
Defines the grounding of the measuring cell. Select one of the follow-
ing options:
Reference
Grounding is ensured by a reference electrode - general use.
Metal
Grounding is ensured by a metal electrode connected to the GND
socket on the Titration Manager. Use this option in case of high resis-
tive solutions in order to avoid measuring background noise at the
electrodes.
Others
Grounding is not ensured by the reference electrode and is defined
outside the method.
Enter in:
Edit method menu
Cell window
Use LEFT/RIGHT arrow keys to access.
This window controls the stirring function of the measurement cell.
An external stirrer (propeller) can be connected to a Titration Manager
or an ABU52.
.
Change
electrode name
1.
Display the Electrode window.
2.
Press 4 then 2.
3.
In the ID field, enter the new name for the electrode (16 charac-
ters maximum).
Start/stop stirrer
Select stirring speed: 100 to 1100 rpm
Animated icon: indicates when stirrer
or propeller is in operation
Select the instrument having the stirrer
(TIM for Titration Manager, ABU1 or ABU2
for an ABU52).