Standard menu, Standard unit incompatible, Titra lab – Hach-Lange TITRALAB 860_865 Reference Manual User Manual
Page 212

Titra
Lab
®
TitraLab 860 and 865 Reference Manual
Page 212
Standard menu
This menu is available for reagent that are edited with the
Titre = Calibrate option. In this menu, identify and enter
the concentration of the standard solution that will be used for the
reagent calibration.
To access:
1.
From the Reagent window, press 4.
2.
Select the reagent to be edited.
3.
Press 2 and check that the Titre = Calibrate option has
been selected.
4.
Use the LEFT/RIGHT arrow keys to move to the last display.
5.
Press 2 Standard.
Standard unit
incompatible
Manual reprocessing error: The calibration method used for the
reprocessing has a standard definition different from the calibration
curve stored.
Edit the parameters of the calibration method to be reprocessed
(Standard menu).