TitraLab 860 and 865 Reference Manual

Page 212

Standard menu

This menu is available for reagent that are edited with the
Titre = Calibrate option. In this menu, identify and enter
the concentration of the standard solution that will be used for the
reagent calibration.

To access:

1.

From the Reagent window, press 4.

2.

Select the reagent to be edited.

3.

Press 2 and check that the Titre = Calibrate option has
been selected.

4.

Use the LEFT/RIGHT arrow keys to move to the last display.

5.

Press 2 Standard.

Standard unit
incompatible

Manual reprocessing error: The calibration method used for the
reprocessing has a standard definition different from the calibration
curve stored.

Edit the parameters of the calibration method to be reprocessed