Page 83

Create reagent

1.

Enter the Reagents window.

2.

Press 4 then 1.

3.

Press

✓ in the New reagent menu.

4.

Select From = Other.

The option From = Catalogue allows you to create a reagent
from a list of commonly used reagents.

5.

In the ID field, enter the reagent name (up to 14 alphanumeric
characters). It is advised to enter the chemical formula of the rea-
gent followed by its concentration (e.g. HCl 0.1).

6.

Enter the "approximate" titre of the reagent (5 characters) in the
Target titre field.

7.

Enter the unit indicated on the reagent bottle (mM = mmol/l, M =
mol/l, mN = meq/l or N = eq/l).

8.

Press 1 twice to create the reagent.

9.

Enter the reagent parameters.

Current value

This is the current sent to the Pt-Pt socket on the titration system
(Titration Manager or ABU52). This parameter is available if
mV(i>0) has been selected for Measurement in the Edit
method or Edit reagent menus.

Enter in:

Method parameters menu
Reagent parameters menu

Range available:

-1000 to +1000 µA in steps of 1 µA

Curve

Select if you want to print the pH/mV = f(volume) titration curve at the
end of each test.

Enter in:

Edit method > Printouts
Edit reagent > Printouts

Curves data lost
- Cal. Data kept -
Methods kept

The last curve data acquisition is lost. Generally, this error occurs
when the instrument is switched off while an analysis is in progress.