Beaker menu, Titra lab – Hach-Lange TITRALAB 860_865 Reference Manual User Manual

Page 67

Advertising
background image

Titra

Lab

®

TitraLab 860 and 865 Reference Manual

Page 67

Beaker menu

Use this menu to prepare a sample or calibration stack. It defines
individual data for all samples/standards used in the sequence.

Figure 11: Beaker menu (for a sample stack)

To access (for a sample stack):

1.

Select Working mode = Sequence or SAC Sequence in the Main
window.

To get access to a SAC sequence, a Sample Changer must
be declared in the Configuration menu.

Refer to "Sample

changer", page 198

.

2.

Press 2 Sequence/Sample stack.

3.

Press 1 Sample stack.

The sequence must have been edited in the Edit sequence
menu.

Refer to "Edit sequence menu", page 97

.

Refer to "Sample stack", page 202

.

To access (for a reagent calibration stack):

1.

Select Working mode = SAC Sequence in the Main window.

2.

In the Reagent window, press 1 Calibrate/Enter titre.

3.

Press 3 Calibration sequence.

The reagent calibration method must have been edited.

Refer to "Edit reagent menu", page 96

.

Refer to "Reagent calibration stack", page 171

.

To access (for an electrode calibration stack):

1.

Select Working mode = SAC Sequence in the Main window.

2.

In the Electrode window, press 1 Calibrate electrodes

3.

Press 2 Calibration sequence.

The electrode calibration method must have been edited.

Refer to "Edit electrode menu", page 94

.

Refer to "Electrode calibration stack", page 101

.