TitraLab 860 and 865 Reference Manual

Page 220

Temperature
Probe/ Fixed at
25°C/Entered

Probe

The sample or standard temperature can be measured using a tem-
perature sensor connected to the Temp socket of the Titration Man-
ager. Select a temperature sensor in the Method parameters menu.

Fixed at 25°C

If measurements are to be performed at a constant temperature of
25°C.

Entered

If the temperature is to be entered manually.

Select in:

Edit method menu
Edit reagent menu if Titre = Calibrate
Edit electrode menu if Calibrate request = Yes

Temperature
sensor ID

Name of the temperature sensor or name of the electrode with a built-
in temperature sensor.

Enter in:

Edit Electrode > Calibration parameters menu
Edit Method > Method parameters
Edit reagent > Reagent parameters

Range available:

16 characters

Test amount

Sample quantity to be placed into the beaker.

Enter in:

Menu Beaker (Menu Sequence/Sample stack then press 1 Sample
stack).

Refer to "Sample stack", page 202

.

Range available:

0,001 to 100000 (unit = Sample Unit)

The sequence is
empty

You are trying to run or complete a sample stack/sequence experi-
ment which contains no programmed methods.

Go to Sequence/Sample stack, Edit sequence screen and check the
sequence.