Select reagent, Select sequence, Sensitivity – Hach-Lange TITRALAB 860_865 Reference Manual User Manual

Page 204: Titra lab

Advertising
background image

Titra

Lab

®

TitraLab 860 and 865 Reference Manual

Page 204

Select reagent

Routine user

Is able to select a reagent to check the parameters and/or start a cali-
bration.

1.

Select the method or the sequence using the reagent.

2.

Press 1 to start a calibration or 3 to check the reagent parame-
ters. In both cases, the available reagents are those specified in
the actual working method or sequence.

Supervisor

From the Reagent window.

1.

Press 4.

2.

In the ID field, select the reagent from the User list.

Select
sequence

To select a sequence (sample changer not in use):

1.

Select Working mode = Sequence in the Main window.

2.

Press 2 Sequence/Sample stack.

To select a sequence to be run using a Sample Changer:

1.

In the Configuration menu, define the model of sample changer
used.

Refer to "Sample changer", page 198

.

2.

Select Working mode = SAC Sequence in the Main
window.

3.

Press 2 Sequence/Sample stack.

Refer to "Sequence/SAC sequence", page 205

.

Sensitivity

Measure of the electrode condition. For ideal electrode chains the
sensitivity is 100%. Generally, however, it is lower. It is expressed as a
percentage of the theoretical slope (59.16 mV/pH) of the curve at
25°C and is determined during a calibration on at least 2 points.

Advertising