TitraLab 860 and 865 Reference Manual

Page 130

Man. stop
(result
indicator)

Refer to "Result indicators", page 185

.

Manual dosing

A manual dosing consists in adding varying size of reagent volumes
at varying addition speeds with or without measuring a signal. No
measurements are stored.

1.

Select the method,

see "Select method", page 203

.

2.

Install the burette,

see "Install burette", page 125

.

3.

Install the reagent on the burette,

see "Install reagent", page 126

.

4.

Connect the electrodes if measurements are to be performed,

see "Electrode connection", page 102

.

5.

If you want to measure pH, run an electrode calibration when a
Stormy icon appears in the Electrode tab.

6.

Press 5 Manual dosing in the Reagent window.

7.

Press 1 Run.

Refer to "Running a manual dosing", page 192

.

Max. stab
reached

Unstable measurement. Stability has not been reached before the
preset Max. Stab time.

Resume the test or end the analysis.

Select the reagent in the list.

Enter the size of increments (0.01 ml to 10 ml).
This volume can be changed during dosing.

Enter the addition speed of increments

(0.01 to 50.00 ml/min).

Enter the maximum volume to dose. Dosing
stops if this volume (0.01 to 1000 ml) is
reached.

Select the electrode in the list.

Select No if you do not want to perform measurements. Otherwise, select
the measurement mode (pH, mV, mV (i>0), Temp.). The available modes

In mV (i>0) mode, select for Applied signalthe current type

Titration Manager or ABU52. The AC signal frequency is 1.67 Hz.

Enter the current value to be applied (from -1000 to +1000 µA in steps of 1 µA).

(alternative AC or direct DC) to be applied to the Pt-Pt socket of the

depend on the electrode type and function selected.