Running a manual dosing, Titra lab – Hach-Lange TITRALAB 860_865 Reference Manual User Manual
Page 192

Titra
Lab
®
TitraLab 860 and 865 Reference Manual
Page 192
Running a
manual dosing
To prepare a manual dosing,
.
Automatic add. = No
An increment is added each time key 1 is pressed.
Automatic add. = Yes
The reagent is added automatically if you press 1. Dosing stops when
the Max. volume has been delivered or if you press 1 again.
Displaying the direct or derivative curve: press Del.
Condition: Measurements must have been performed
Stopping a manual dosing: press Stop.
The Reagent window is displayed.
Adjust the increment size (0.01 ml to 10 ml).
Select Yes to add the increments continuously
Press 1 to start dosing the reagent
(one increment is added or reagent is added
automatically increment by increment ).
Select No to add a given number of
increments.
If measurements have been selected, pH
(or mV) and/or temperature are displayed live.
With measurements: the measured
signal and/or temperature are also
= f(volume) is displayed live.
displayed. Curve measured signal
If no measurements: the volume
already dispensed is displayed.
Curve volume (in ml) = f(time)
is displayed live.
Select the curve to display.
Press 1 to quit the manual dosing.
Use the RIGHT/LEFT arrow keys to move
along the curve and display the point
coordinates shown by the cursor.