Titre enter/ calibrate, Tray missing (sac error), Ttl 5 v out/ ttl 12 v out (sockets) – Hach-Lange TITRALAB 860_865 Reference Manual User Manual
Page 222: Titre enter/calibrate, Ttl 5 v out/ttl 12 v out (sockets), Titra lab

Titra
Lab
®
TitraLab 860 and 865 Reference Manual
Page 222
Titre Enter/
Calibrate
Enter
If you already know the reagent concentration and do not wish to per-
form a calibration.
Calibrate
If you wish to perform a calibration. The Titration Manager will prompt
you to enter the calibration parameters.
It is not necessary to calibrate an addition reagent. It is there-
fore recommended to select Titre = Enter so that the calibration
parameters are not displayed.
Enter in:
Edit reagent menu.
Tray missing
(SAC error)
On starting a sequence, the sample changer cannot read the RFID
tag of the turntable. Check that a turntable is correctly mounted on the
sample changer.
Check the model of turntable used and refer to the User’s Guide of the
sample changer, chapter 6 "Accessories".
In particular, note that a SAC850 turntable can be used on a SAC950
but a SAC950 turntable cannot be used on a SAC850.
Solve the problem then restart the sequence.
TTL 5 V OUT/
TTL 12 V OUT
(sockets)
Red and black banana sockets (diameter = 2 mm).
The red banana sockets send a TTL 0 ± 5 V or 0 ±12 V.
The black banana sockets are connected to the electrical zero of the
instrument.
This auxiliary signal is programmed in the Method parameter or Rea-
gent parameter menus,
see "Auxiliary output", page 56
.
Specifications (potential difference is given in absolute value)
Output signal
Level 0: 0 to 0.4 V
Level 1: higher than 2.4 V and equal to or less than 5 V (or 12 V)
Output impedance: 1 KOhm at level 0 and 2 kOhm at level 1
A TTL 0-level does not mean a 0 V output. The voltage differ-
ence varies from 0 to ±400 mV.