Ph int, Titra lab – Hach-Lange TITRALAB 860_865 Reference Manual User Manual
Page 150
Advertising
Titra
Lab
®
TitraLab 860 and 865 Reference Manual
Page 150
pH int
This parameter is available when creating a single or a combined pH
electrode with the option From = Other.
Refer to "Create electrode", page 82
This is the internal pH of a single or combined pH electrode.
The pH int is used for the buffer recognition.
The table below gives the pH int of Radiometer Analytical pH
electrodes:
Table 7: pH int of Radiometer Analytical electrodes
Range available:
0.00 to 14.00 pH
Electrode
pH int
PHG301, PHG311, XG100, XG200, XG250
6.06 pH
PHG201, PHG311
6.65 pH
PHC2001, PHC2011, PHC2085,PHC2401,
PHC2501, PHC2601, PHC3001, PHC3011,
PHC3081, PHC3185, XC100, XC111, XC120,
XC161
6.65 pH
PHC4000
6.80 pH