Number of digits, Number of dynamic rinses, Number of ep – Hach-Lange TITRALAB 860_865 Reference Manual User Manual
Page 144: Number of equations, Titra lab

Titra
Lab
®
TitraLab 860 and 865 Reference Manual
Page 144
Number of
digits
Number of significant digits (1 to 5) to be displayed and printed for the
result calculated.
Example: If Number of digits = 4:
•
1456.1 is displayed “1456”
•
12.124 is displayed “12.12”
•
0.15872 is displayed “0.1587” (the first significant digit is “1”)
•
0.4 is displayed “0.4000” (the first significant digit is “4”)
Enter in:
Edit method > Results (for titration methods)
Edit reagent > Results
Number of
dynamic rinses
If a SAC850 or SAC950 Sample Changer is in use, enter the desired
number of dynamic rinses to be carried out before each beaker
analysis of a SAC sequence,
see also "Dynamic rinses", page 93
.
Enter in:
Setup menu > Configuration
Range available:
0 to 9
Number of EP
Number of end points (1 to 4) to be titrated to.
Only one end point is obtained during a back titration or a
reagent calibration.
Enter in:
Edit method > Method parameters menu (for End point method)
Number of
equations
This is the number of extra results (0 to 2) to be calculated by the
method. These results are calculated using a user-programmed equa-
tion. This allows you to calculate the result in units which are not avail-
able in the instrument.
Enter in:
Edit method > Results