Page 143
Number of
additions
Set the number (1 to 3) of additions to be performed automatically by
an Addition method.
Refer to "Addition method - definition", page 49
.
Enter in:
Edit method menu (Addition method)
Number of
buffers
Available if Calibration request = Yes
Number of pH standards to be used for the calibration. Work with at
least two standards to calculate the electrode sensitivity. If one stand-
ard is only used the zero pH is calculated.
Enter in:
Edit electrode menu
Range available:
1 to 5
Refer to "pH calibration solutions parameters", page 149
.
Number of
cycles
Available if Calibration request = Yes
Number of times the calibration is to be repeated, i.e. the number of
beakers to be prepared for each pH standard.
Enter in:
Edit electrode menu
Range available:
1 to 9
Number of
decimals
Number of decimals (0 to 3) to be displayed and printed for the result.
Enter in:
Edit method > Results (for a pH measurement method)