Number of additions, Number of buffers, Number of cycles – Hach-Lange TITRALAB 860_865 Reference Manual User Manual

Page 143: Number of decimals, Titra lab

Advertising
background image

Titra

Lab

®

TitraLab 860 and 865 Reference Manual

Page 143

Number of
additions

Set the number (1 to 3) of additions to be performed automatically by
an Addition method.

Refer to "Addition method - definition", page 49

.

Enter in:

Edit method menu (Addition method)

Number of
buffers

Available if Calibration request = Yes

Number of pH standards to be used for the calibration. Work with at
least two standards to calculate the electrode sensitivity. If one stand-
ard is only used the zero pH is calculated.

Enter in:

Edit electrode menu

Range available:

1 to 5

Refer to "pH calibration solutions parameters", page 149

.

Number of
cycles

Available if Calibration request = Yes

Number of times the calibration is to be repeated, i.e. the number of
beakers to be prepared for each pH standard.

Enter in:

Edit electrode menu

Range available:

1 to 9

Number of
decimals

Number of decimals (0 to 3) to be displayed and printed for the result.

Enter in:

Edit method > Results (for a pH measurement method)

Advertising