Stirring, Titra lab – Hach-Lange TITRALAB 860_865 Reference Manual User Manual
Page 215

Titra
Lab
®
TitraLab 860 and 865 Reference Manual
Page 215
Stirring
Internal stirrer, e.g. magnetic
Display the Cell window.
.
External stirrer, e.g. propeller
1.
Connect the Stirring Propeller, part no. 904-543, to the Propeller
socket.
2.
Display the Cell window.
3.
Select External stirring = On and adjust stirring by
turning the stirrer propeller knob or Press
✓ in the Speed
setting field and select a stirring speed from the table. .
4.
To stop stirring, select External stirring = Off.
SAC850 and SAC950 stirrer: the stirring speed is adjusted
using the TitraMaster 85 PC software.
Start/stop stirrer
Select stirring speed: 100 to 1100 rpm
Animated icon: indicates when stirrer
or propeller is in operation
Select the instrument having the stirrer
(TIM for Titration Manager, ABU1 or ABU2
for an ABU52).
Start/stop stirrer
Select the instrument the stirrer is connected
to (TIM for Titration Manager, ABU1 or ABU2
for an ABU52).