Reprocessing a curve, Titra lab – Hach-Lange TITRALAB 860_865 Reference Manual User Manual

Page 179

Advertising
background image

Titra

Lab

®

TitraLab 860 and 865 Reference Manual

Page 179

Reprocessing a
curve

This method mode enables to reprocess the last curve obtained using
another method or method parameters that differs from the original
ones (parameters defined before obtaining the curve). Thus, you can
correct or obtain a result.

You can reprocess a curve in 2 ways:

In AUTOMATIC mode using a Coupled method. A so-called
“Reprocess“ method follows next the titration method which
generates the curve being reprocessed.

In MANUAL mode by selecting Working mode =
Reprocessing in the Main window. Modify the method
parameters (if requested) then run the method.

The last obtained reagent calibration curve can also be reprocessed
in MANUAL mode. Change the calibration method parameters then
run the Calibration clicking the “Reprocessing“ command from the
Calibrate/Enter titre window.

In MANUAL or AUTOMATIC mode, you can reprocess a reprocessed
curve.

You can only reprocess an End Point curve in End Point mode.

Advertising