Equivalent point determination (ip methods), Titra lab – Hach-Lange TITRALAB 860_865 Reference Manual User Manual

Page 113

Advertising
background image

Titra

Lab

®

TitraLab 860 and 865 Reference Manual

Page 113

Equivalent point
determination
(IP methods)

An equivalent point of a titrated species/titrant reaction is determined
in two ways depending on the titrant addition mode selected:

Continuous addition of the titrant (Continuous IP method):

•

acquisition of the titration curve E (electrode) = f (volume of titrant
dispensed),

•

calculation of the first derivative curve with smoothing process
(method parameter: smoothing criterion),

•

determination of the second derivative curve,

•

determination of the inflection points on the first derivative curve
from the minimum and maximum of the second derivative curve,

•

determination of the tangents to the first derivative curve for these
two inflection points,

•

determination of the equivalent point as the intersection of the two
tangents.

Figure 19: Determination of an equivalent point - Continuous IP method

An equivalent point is determined and plotted live on the titration
curve. The method can be stopped after a given number of equivalent
points has been detected.

E or pH