Equivalent point determination (ip methods), Titra lab – Hach-Lange TITRALAB 860_865 Reference Manual User Manual
Page 113

Titra
Lab
®
TitraLab 860 and 865 Reference Manual
Page 113
Equivalent point
determination
(IP methods)
An equivalent point of a titrated species/titrant reaction is determined
in two ways depending on the titrant addition mode selected:
Continuous addition of the titrant (Continuous IP method):
•
acquisition of the titration curve E (electrode) = f (volume of titrant
dispensed),
•
calculation of the first derivative curve with smoothing process
(method parameter: smoothing criterion),
•
determination of the second derivative curve,
•
determination of the inflection points on the first derivative curve
from the minimum and maximum of the second derivative curve,
•
determination of the tangents to the first derivative curve for these
two inflection points,
•
determination of the equivalent point as the intersection of the two
tangents.
Figure 19: Determination of an equivalent point - Continuous IP method
An equivalent point is determined and plotted live on the titration
curve. The method can be stopped after a given number of equivalent
points has been detected.
E or pH