Simultaneous
additions (Yes/
No)

In an Addition method, you can run automatically 2 or 3 additions.
These additions can be performed simultaneously or one after the
other.

Refer to "Addition method - definition", page 49

.

The reagents are added using burettes all controlled by the Titration
Manager. These burettes can be installed on the Titration Manager or
on an ABU52 connected to the Titration Manager.

Enter in:

Edit method menu (Addition method with 2 or 3 additions).

Skip empty
position

Parameter of a SAC Sequence for sample beaker positions only. Does
not apply to electrode (or reagent) calibration beaker positions.
Does not apply if a SAC850 or SAC950 sample changer is used with
the Beaker detection option cleared (deselected),

see "Beaker detection", page 65

.

Selects whether you want or not the sample changer (SAC80, SAC90,
SAC850 or SAC950) to skip to the next beaker position if an empty
position (*) is found. If No is answered, 3 possibilities are offered if an
empty position is found: restart the analysis in the same beaker, skip
to the next beaker or end the analysis.

Enter in:

Press 2 Sequence/Sample stack from the Main window.

Refer to "Programming sequence", page 159

.

(*) Empty position:
An empty position means no beaker present at the position.
When you are using a SAC850 or SAC950, an empty position also
means a beaker with less liquid than the minimum detection limit or a
beaker with a solid or a powder sample.

Refer to "Beaker detection minimum height", page 66

.