Beaker detection minimum height, Titra lab – Hach-Lange TITRALAB 860_865 Reference Manual User Manual
Page 66

Titra
Lab
®
TitraLab 860 and 865 Reference Manual
Page 66
Beaker
detection
minimum height
The table below reports the minimum height of liquid that must be
present in the beaker in order to detect the beaker as not empty,
also "Beaker detection", page 65
.
Beaker type
diameter x height (mm)
Detection minimum
height
Part no.
400 ml tall form 70 x 131
10 mm (40 ml)
250 ml low form 70 x 95
10 mm (40 ml)
250 ml tall form 60 x 120
10 mm (30 ml)
125 ml, Gosselin, PP,
54 x 73
10 mm (25 ml)
X31T087 (pack of 50)
180 ml, Gosselin, PP,
54 x 103
10 mm (25 ml)
X31V005 (pack of 50)
150 ml tall form 60 x 80
10 mm (30 ml)
150 ml low form 62 x 81
10 mm (30 ml)
40-100 ml, PP 60 x 80
10 mm (30 ml)
904-490 (pack of 50)
100 ml tall form 48 x 80
10 mm (20 ml)
50 ml low form 42 x 60
10 mm (15 ml)
904-489 (pack of 50)
22-45 ml, PP 44 x 70
10 mm (15 ml)
904-489 (pack of 50)
8-25 ml, PP 44 x 70
10 mm (15 ml)
904-488 (pack of 50)