Results menu, Rinse aux. output, Titra lab – Hach-Lange TITRALAB 860_865 Reference Manual User Manual
Page 188

Titra
Lab
®
TitraLab 860 and 865 Reference Manual
Page 188
Results menu
This menu displays the result identification parameters and
acceptance criteria required to run a sample method, an electrode or
a reagent calibration.
To access:
•
Method: Method library > Edit method > Results (key 3).
•
Reagents: Reagents library > Edit reagent > Results (key 3).
•
Electrodes: Electrode library > Edit electrode > Results (key 3).
The Results menu for a sample method with 1 result and 1 equation
Rinse aux.
output
Option available when sample changer in use.
This command will set the auxiliary signal on during a programmed
rinse. For example, to clean the gas diffusion tip when using the
FRB560 Gas Flow Regulation Kit.
Enter in:
Setup menu > Configuration
Range available:
No, 5 V, or 12 V