Reprocessing a curve (automatic mode), Titra lab – Hach-Lange TITRALAB 860_865 Reference Manual User Manual

Page 181

Advertising
background image

Titra

Lab

®

TitraLab 860 and 865 Reference Manual

Page 181

Reprocessing a
curve
(Automatic
mode)

Perform the following operations to reprocess a curve in
automatic mode (Supervisor mode only):

1.

Create a Coupled method,

see "Programming methods", page

29

.

2.

For line 1(or at the line preceding the so-called “Reprocess“
method): select the method which generates the curve.

3.

Select the so-called “Reprocess“ method at the next line.

(*)

Refer to "Reprocessing a curve (rules)", page 180

.

4.

Select Yes for Use for reprocessing.

Run the Coupled method to start the curve reprocessing

The reprocessed results are stored with the “Reprocessed“ indication.
To view results,

see "GLP-Archives menu", page 122

.

An EP or IP method must be selected.

This method must be edited with specific
rules (*).

Then press

to confirm.

A method selected with the Reprocessing option
is displayed with its number in brackets in the
sequence.
For example:
1 Alkalinity
(2) Alkalinity-Rep

Advertising