Reagent
calibration not
required

Message appears at the start of a sequence, when a method
sequence had been programmed with a reagent calibration. The rea-
gent has been reprogrammed without a calibration.

Go to Sequence/Sample stack, Edit sequence screen and remove the
reagent calibration from the sequence.

Reagent
calibration
parameters

This menu contains the general parameters concerning the reagent
calibration method (identification of the electrodes used in particular).
Titration parameters necessary to run the calibration are also pro-
grammed.

Do not forget to select the electrode and temperature sensor
created in this menu!

To access:

1.

From the Reagent window, press 4.

2.

Select the reagent to be edited.

3.

Press 2 Edit reagent and check that the Titre = Cali-
brate option has been selected.

4.

Edit the calibration general parameters.

5.

Use the LEFT/RIGHT arrow keys to move to the last Edit reagent
display.

6.

Press 1 Calibration parameters.