Increment, Inflection point (ip), Input address conflict – Hach-Lange TITRALAB 860_865 Reference Manual User Manual
Page 124: Titra lab

Titra
Lab
®
TitraLab 860 and 865 Reference Manual
Page 124
ID
Refer to "Electrode ID", page 105
Refer to "Equation ID", page 111
Refer to "Excess reagent ID", page 117
.
Refer to "Method ID", page 133
.
.
Refer to "Reagent addition ID", page 167
Refer to "Reagent ID", page 173
Refer to "Result ID", page 185
Refer to "Sample ID", page 200
.
Refer to "Standard ID", page 211
.
Refer to "Sequence ID", page 206
.
Refer to "Temperature sensor ID", page 220
.
Refer to "Titrant ID", page 221
Refer to "User ID (Yes/No)", page 224
Increment
Sets the size of the increments of a Monotonic IP titration. The stand-
ard value of 10% of the burette size will be suitable for most applica-
tions. For titrations where the inflection on the titration curve is very
abrupt (typically redox titrations), use a smaller value. Decreasing the
Increment slows down the titration.
The Increment is set in the Parameters menu of a Monotonic IP
method.
Range available:
0.01 ml to 10 ml.
Inflection Point
(IP)
Corresponds to the inflection points of the E (electrode) = f (volume of
titrant dispensed) titration curve. These inflection points represent the
equivalent points of the titrated species/titrant reaction. An inflection
on the titration curve is shown as a peak on the derivative curve. With
the Titration Manager, up to eight inflection points can be located and
used for result calculation during a single titration.
Input address
conflict
Two electrodes have been defined at the same address.
Enter the Edit electrode menu and modify the address of one of the
electrodes.