Copy method

1.

Switch to Main window.

2.

Select Method.

3.

Press 4 then 1.

4.

Press 3 in New method menu.

5.

Enter a method name.

6.

Press 2 Copy from method.

7.

In the Library field, select the Preprogrammed or User list.

8.

In the ID field, select the method to be copied from the list of avail-
able methods.

9.

Press 1 to confirm. The method is created and saved in the User
list.

Copy reagent

This procedure is used to create a reagent (new reagent ID), without
having to enter data nor program the corresponding calibration
method.

1.

Enter the Reagents window.

2.

Press 4 then 1.

3.

Press

✓ in the New reagent menu.

4.

Select From = Catalogue.

5.

In the ID field, select a reagent name from the Catalogue list.

6.

Enter the "approximate" titre for the reagent (between 0.001 and
99 999) in the Target titre field.

7.

Enter the unit indicated on the reagent bottle (mM = mmol/l,
M = mol/l, mN = meq/l or N = eq/l).

8.

Press 1 to confirm then 2 Copy from reagent.

9.

In the Library field, select the Preprogrammed or User list.

10. In the ID field, select the reagent to be copied from the list of

available reagents.

11. Press 1 to confirm. The reagent is created and saved in the User

list.

It is not necessary to select Catalogue to create a rea-
gent using the copy function. A reagent ID created from
Other can also use the copy function.