Equivalent point determination (ip methods), Titra lab – Hach-Lange TITRALAB 860_865 Reference Manual User Manual

Page 114

Advertising
background image

Titra

Lab

®

TitraLab 860 and 865 Reference Manual

Page 114

Equivalent point
determination
(IP methods)

Incremental addition of the titrant (Monotonic/Dynamic IP
methods):

•

acquisition of the titration curve E (electrode) = f (volume of titrant
dispensed),

•

determination of the first derivative curve using a smoothing proc-
ess (depending on the value set for IP filter),

•

determination of the second derivative curve,

•

determination of the maxima of the first derivative curve using the
IP filter parameter. These maxima correspond to the zeros of the
second derivative curve.

•

determination of the equivalent point: for this point the maximum of
the first derivative curve must be equal to or greater than the preset
IP reject method parameter.

Figure 20: Determination of an equivalent point - Monotonic or Dynamic IP

method

An equivalent point is determined and plotted live on the titration
curve. The method is stopped after a given number of equivalent
points has been detected.

E or pH