Create electrode, Create method, Titra lab – Hach-Lange TITRALAB 860_865 Reference Manual User Manual
Page 82

Titra
Lab
®
TitraLab 860 and 865 Reference Manual
Page 82
Create electrode
1.
Enter the Electrode window.
2.
Press 4 then 1.
3.
Select the electrode function,
see "Electrode type", page 106
.
4.
Press
✓ in the ID field.
5.
Select From = Other.
The option From = Catalogue allows you to create an
electrode from a list of Radiometer Analytical electrodes.
6.
Enter the electrode name (up to 16 alphanumeric characters).
7.
In the Confirm creation screen:
•
For electrodes with pH or mV functions only; select the electrode
type.
Refer to "Electrode type", page 106
.
•
For combined pH or single pH electrodes; enter the internal pH
(pH int) of the electrode.
.
•
For combined pH, Metal/Redox or ISE electrodes, select if the
electrode has a built-in temperature sensor or not.
•
For reference electrodes or combined pH, Metal/Redox or ISE
electrodes, enter the potential of the reference (in mV) versus the
Standard Hydrogen Electrode.
Refer to "Potential versus SHE", page 151
8.
Press 1 to create the electrode. The Edit electrode menu is dis-
played. Enter the electrode definition parameters.
Create method
1.
Enter the Main window.
2.
Select Working mode = Method.
3.
Press 4 then 1.
4.
Press
✓ in the New method screen.
5.
Enter a method name (up to 17 alphanumerical characters).
6.
Press 1 to create the method.
Go to Edit method screen and enter the method parameters.