Reprocessing a
curve (Manual
mode)

Perform the following operations to reprocess a curve in manual
mode (Supervisor mode only):

1.

Create then run the method that generates the curve being
reprocessed. An EP or IP method must be used.

2.

Select the method you want to use in reprocessing mode.

Refer to "Select method", page 203

.

3.

In the Main window, select Working mode = Reprocessing.

4.

You can change the method parameters. This method must be
edited according to specific rules,

see "Reprocessing a curve

(rules)", page 180

.

5.

Run the method by pressing 1 from the Main window.
At the end of the reprocess, a result screen is displayed with 3
possible options:

The reprocessed results are stored with the “Reprocessed“
indication. To view results,

see "GLP-Archives menu", page 122

.

The Manual mode can be used to adjust parameters like the IP reject,
IP filter or Smoothing parameter.

Note that in Reprocessing mode,
calculator icons are displayed in the
Reagent and Electrode tabs.

To save the results and quit.

To quit without saving the results.

To display result acceptation screens,