Start addition

Option available with TitraLab 865 or TitraLab860 with one or two
ABU52 connected.

Define when the reagent addition will be performed.

Start addition = Initial addition

The addition is performed at the start of the titration (titrant predose
included).

Start addition = After EP no. x (or IP no. x)

The addition is performed after EP no. x (or IP no. x). The titration to
the next EP (or IP) starts at the end of the time entered for D
elay after
addition.

Enter in:

Edit method > Parameters (if Addition = Yes)
Edit reagent > Calibration parameters (if Addition = Yes)

Initial addition is the only option available in Continuous IP
mode.

Start from

Command of the Sequence/Sample stack window that enables to
start a TIM or SAC sequence from a beaker number of your choice.

Refer to "Sequence/Sample stack menu", page 206

.

Refer to "Running a TIM sequence", page 195

.

Refer to "Running a SAC sequence", page 194

.

Start timer

This is the time delay the Titration Manager will respect before starting
the titration. If a predose is programmed, this delay will be counted
down at the end of the predose. The Start timer delay will allow the
solution to stabilise before performing the first measurement.

While an addition method is running, you can carry on
regardless this delay by pressing key Del
.

Enter in:

Edit method > Parameters menu
Edit reagent > Calibration parameters menu

Range available:

00:00 to 99:59 min:s