Addition volume, Address, Alarm: locked – Hach-Lange TITRALAB 860_865 Reference Manual User Manual

Page 51: Titra lab

Advertising
background image

Titra

Lab

®

TitraLab 860 and 865 Reference Manual

Page 51

Addition volume

Refer to "Reagent addition volume", page 167

.

Address

The position where the electrode and burette are placed during opera-
tion:

Electrode

The electrode address is defined using the format “xxx/y” where “xxx”
corresponds to the instrument (TIM, ABU1 or ABU2) where the
electrode is connected and “y” corresponds to the socket.
For example TIM/E1, indicates that the electrode is connected to E1
socket on the Titration Manager.

It is always recommended to connect an indicating electrode
and its associated reference to the same instrument (TIM,
ABU1 or ABU2).

Refer to "Electrode connection", page 102

.

Burette

The burette address is defined using the format “xxx/y” where “xxx”
corresponds to the instrument (TIM, ABU1 or ABU2) where the burette
is placed and “y” corresponds to the position. For example TIM/1, indi-
cates that the burette is placed on the Titration Manager in position 1.

Figure 2: Burette positions

Refer to "ABU1/ABU2", page 43

.

Alarm: Locked

The user cannot bypass an electrode, reagent calibration and/or QC
sample analysis if the last result obtained lies outside the acceptance
range.

Enter in:

Setup menu > Access routine mode

Position 2

Position 1