Recalculate
results
You can recalculate a result at the end of a sample titration or reagent
calibration method. The old result is replaced with the new. The
recalculated result is saved with the indication "Recalculated".
Proceed as follows to recalculate a result:
•
At end of titration, press 2 More details the 3 Recaculate in the
results screen.
•
Change the calculation parameters.
•
Recalculated results are displayed.
You can accept then save the recalculated results (press 1 Back then
1 Save and continue). To view the results,
A result that has been accepted by the instrument can be
rejected after recalculation (and vice versa).
The test quantity can be modified for each
test performed. If pre-diluted samples are
used, the Sample amount and Final dilution
amount can also be changed for the first
test of each sample (or analysis).
Press 2 to recalculate the result(s) or 1 to
return to the first screen without
recalculating the result(s).
You can start a new recalculation (press 2)
or return to the first screen above (press 1).