Titra lab, Environmental operating conditions, International standards – Hach-Lange TITRALAB 860_865 Reference Manual User Manual
Page 244

Titra
Lab
®
TitraLab 860 and 865 Reference Manual
Page 244
Environmental operating conditions:
Interior use only
Ambient temperature:
- working range: 5 to 40°C
- storage: -20 to 60°C
- transport: -40 to 60°C
Relative humidity:
20 to 80 % with temperature between 5 and 31°C. Above 31°C, the interval amplitude
decreases linearly from 20 - 80% at 31°C to 20 - 50% at 40°C.
Altitude
≤ 2000 m.
Level of pollution : 2
Transitory overvoltage: class II.
International standards
Refer to "International Standards", page 232.
Burette specifications according to ISO 8655-3
Burette
Nominal
Maximum permissible Maximum permissible
stand
volume
systematic errors
random errors
Type
ml
± %
± µl (a)
± % (b)
± µl (c)
B501
< 1
0.6
6
0.1
1
B505
5
0.3
15
0.1
5
B510
10
0.2
20
0.07
7
B525
25
0.2
50
0.07
17.5
B550
50
0.2
100
0.055
25
(a)
Expressed as the deviation of the mean of a tenfold measurement from the nominal volume or from
the selected volume (see ISO 8655-6:—, 8.4).
(b)
Expressed as the coefficient of variation of a tenfold measurement (see ISO 8655-6:—, 8.5).
(c)
Expressed as the repeatability standard deviation of a tenfold measurement (see ISO 8655-6:—, 8.5).