No curve stored within the instrument, No stirrer (sac error), Notification message – Hach-Lange TITRALAB 860_865 Reference Manual User Manual
Page 142: Titra lab

Titra
Lab
®
TitraLab 860 and 865 Reference Manual
Page 142
No curve stored
within the
instrument
Manual reprocessing error: you are trying to reprocess in manual
mode a curve and there is no curve stored.
Reprocessing is not possible as a curve has to be stored in the
instrument.
Refer to "Reprocessing a curve (Manual mode)", page 182
No stirrer (SAC
error)
SAC950 Sample changer error: there is no magnetic stirring possible
for the beaker position mentionned.
When using a 2 or 3-radii turntable with a SAC950, you must use a
propeller stirrer to stir the solutions that are placed on the smallest
row (2-radii turntable) or on the 2 smallest rows (3-radii turntable).
Connect a propeller stirrer to the sample changer. Refer to the Sam-
ple Changer User’s Guide, part no. D21T085.
You can also edit a sample stack in order to use a 1-radius turntable
or to use only the largest row of a mutil-radii turntable. In this case, the
magnetic stirring is possible.
Refer to "Sample stack", page 202
Notification
message
Select Notification = Yes if you want to a message to be
displayed on starting a reagent calibration, a measurement or titration
method. Type the message (3 lines of 1 to 32 alphanumerical charac-
ters).
Enter in:
Edit method menu
Edit reagent menu
Green positions : magnetic and propeller stirring possible
Red positions : propeller stirring only