Reagent system
A reagent system comprises all the reagents necessary to run a
method or a sequence of methods:
When a method/sequence is run, the instrument prompts you
to install or replace the reagents that will be required to run this
method/sequence.
Reagent titre
not entered
The reagent titre of the active reagent has not been entered in the
Reagent library.
Press 1 in the Reagents window, press 1 Enter titre.
Reagent unit
Specify the reagent concentration unit. The units are given on the rea-
gent bottle or are the unit used to express the titrant concentration
after calibration.
Enter in:
Reagent library > Edit reagent menu.
Range available:
Select, mM = mmol/l, M = mol/l, mN = meq/l or N = eq/l.
If molar units are selected (mM or M), it is necessary to enter
stoichiometric coefficients for the chemical reaction.
Once the units have been confirmed, they are added to the
reagent ID.
Reagent
window
This window contains all the information and operations concerning
the reagents.
To access:
Use LEFT/RIGHT arrow keys